Oxylos Leak_O2
5 automated tests in 1: O2 gas analysis; micro-leak measurement system; creep test (A&B); packaging sealing resistance. Without consumables
Full Description
5 automated tests in 1: O2 gas analysis; micro-leak measurement system; creep test (A&B); packaging sealing resistance. Without consumables
The OXYLOS LEAK/O2 allows you to measure the rate of O2 and the airtightness across all types of containers and packaging, with protective atmosphere, and without having to use any consumables (no need to use needles or septums). Automated and safe, the OXYLOS can be adapted to your production constraints, providing you with sampling tools and customised positioning. The high-end analyser from the ABISS brand provides you with solutions for packaging when space above the contents is limited. Able to be synchronised to the packaging line, the OXYLOS becomes autonomous and does not require an operator in order to carry out measurements. This analyser is ideal for production.
THE + PRODUCTS
- Meets standards:
DIN 55508-1: measures flow rate at constant pressure
ASTM F1140: seal strength
ASTM F2054: burst test
- The option of combining 6 measurements in 1 same test: O2/CO2 gas analysis; Micro-leaks; Creep test (A&B); Bursting
- No consumable
- Fully hand safe
- Laser positionning tool for sample
- Can be synchronised to conveyors
- Effective hole size calculation down to 5 μm
- Solutions where available space above the product is limited
- Fully traceable via an internal spreadsheet
- The needles are cleaned automatically
- Remote maintenance 7 days a week, with remote connection available
Features
| Custom placement tools | yes |
| Ingress protection of enclosure | IP65 connectivity |
| Measurement device | Bench-mounted |
| Enclosure | Stainless steel |
| Weight | 28 kg |
| Width | 560 mm |
| Height | 495 mm |
| Depth | 473 mm |
| Power | 100-240VAC -50/60Hz - 63W |
| Feed pressure | 4 to 8 Bar (dry and oil free filtred to 20 µm) filtered at 20μm |
| Inlet pneumatic connection diameter | 4 x 6 mm tube connection |
| Memory capacity | 4 to 32go |
| Display | Color touchscreen 7’’ |
| Digital interfaces | 1 USB 2.0 and 1 Ethernet RJ45 |
| Storage conditions | 0 - 70°C/0 - 90% HR |
| Terms of use | 0 - 40°C/10 - 90% HR |
| O2 Oxygen measurement | Technology : Electrochemical Precision : +/- 1% relative T90 response time : |
| Pressure measurement | 1 |
| Leak measurement | Technology : Mass flowmeter Precision : +/- 0.8% full scale (μm) ; +/- 1% of the reading + 0.1 ml/min Resolution : 0.1ml/min - 1μm Measurement : 0.1-500 ml/min - 5 μm to 500 μm Test pressure : From 10 to 500 mbar +/- 1mbar Flow : Filling: 4200 ml/min Compressed air supply : 4 to 8 Bar (dry and oil f |
| Burst test | Technology : Barometric Accuracy : +/- 0,5% full scale Resolution : 1mbar Flow scales : 0,2 – 1,4 bar Quick inflating flow : Minimum 250l/h Burst pressure : Up to 1,4 bar Analytical curves (displayed and recordable) or Creep test Technology : Barometric Accuracy : +/- 0,5% full scale Resolution : 1m |
| Blocked circuit detection | yes |
| Pump suction control | yes |
| Compensation for atmospheric pressure variations | yes |
| Results storage | yes |
| USB Connection | yes |
| Touchscreen | yes |
| Traceability | Format, line, operator, batch number, comment at the end of the test |
| Remote Connection | Remote control possible for remote maintenance |
| Network connection (Ethernet socket) | yes |
| Synchronised automatic triggering | yes |
| Automatic non-septum test | yes |
| Needles automatic cleaning | yes |
| Options | New: Injection gases are continuously monitored Online device synchronised with the conveyor 3G router + HMI Advanced Can interface with ERPs 2nd Ethernet port to isolate the internal from the external network |